PETECOS

UV Fluorescence Sulfur Element Analyzer PTS-3000 Type

PRODUCTS

CONTACT US

UV Fluorescence Sulfur Element Analyzer PTS-3000 Type

Product Description

UV Fluorescence Sulfur Element Analyzer PTS-3000 Type

  1. Overview

The principle of ultraviolet fluorescence measurement is adopted. After the sample is introduced into the high-temperature cracking furnace, the sample undergoes a cracking and oxidation reaction. At a high temperature of about 1000°C, the sample is completely gasified and undergoes oxidative cracking, and the sulfide in it is quantitatively converted into sulfur dioxide. The reaction gas is carried by the carrier gas, passes through the membrane dryer to remove its moisture, and enters the reaction chamber.
Sulfur dioxide is exposed to ultraviolet light of specific wavelengths, and the absorption of this ray causes some electrons to divert to high-energy orbits. Once the electrons return to their original orbits, the excess energy is released in the form of light and is detected and received by a photomultiplier tube at a specific wavelength. The emitted fluorescence is completely specific to sulfur and is proportional to the sulfur content in the original sample.
After amplification by a microcurrent amplifier and computer data processing, it can be converted into an electrical signal proportional to the light intensity. By measuring its size, the sulfur content of the corresponding sample can be calculated.
2.Technical indicators

NO. Project Instrument technical specifications
1 Scope of application It is suitable for measuring the total sulfur content in crude benzene, coal tar, gasoline, diesel, biological grease, lubricating oil, crude oil, distillate oil, liquefied petroleum gas, coal gas, natural gas, plastics, petrochemical products, food, coal, etc.
2 Applicable standards GB/T 11060.8-2012、GB/T 34100-2017、SH/T 0689-2000、GB/T 11141-2014、ASTM D5453-1993、GB/T 35582-2017、GB/T 11060,GB/T17930、ASTM D5453-2016欧洲标准编号EN ISO 20846:2012
3 Measuring range 0.2 mg/L~10000mg/L~ Percent content (liquid)
4 Temperature range room temperature~1100℃
5 Temperature control accuracy ±1℃
6 Working voltage AC220V,3KW
7 Repeatability error (liquid) X≤1mg/L   Cv≤±0.2mg/l
1.0mg/L<X≤10mg/L   Cv≤±10%
10mg/L<X   Cv≤±5%
8 Measurable sample status Solid, liquid, gas (optional injector, standard liquid injection)
9 Workbench 1500mmX600mm
11 Gas source High purity oxygen>99.99%    High purity argon>99.99%

Complete sets of instruments: host (data acquisition, control system), cracking furnace, injector, computer (including operating software), etc.

  1. Performance characteristics
NO. Project Instrument performance characteristics
1 Communication interface The instrument adopts USB communication technology, which makes data collection more accurate and control more accurate
2 Analytical system The instrument truly realizes high-voltage continuous adjustable and over-pressure monitoring, effectively protecting the detection system
3 Baseline The instrument implements baseline automatic zeroing technology
4 Peak shape changes The instrument implements sample peak-row auto-magnification and reduction techniques
5 System Windows operation platform, human-computer dialogue, convenient operation
6 Current design Microcurrent detection adopts the first sulfur detector in China, which makes the instrument have the advantages of high sensitivity, low noise, wide linear range, and strong anti-interference ability, and is an advanced total sulfur content analyzer in China
7 Structural protection The instrument adopts upper and lower split structure, which protects the circuit system and is also convenient for later maintenance and repair
8 Fan control The instrument fan is automatically controlled by the temperature of the cracking furnace, and there is no need to be manned after the analysis is completed
9 Multi-point correction Standard sample correction can be single-point correction or multi-point correction, which is convenient, fast and accurate. (Unique Technology)
10 Components import Using international popular circuits and imported devices, the performance of the whole machine is better than that of similar products in China

 